Dual range pistons for pressure calibration

Dual piston deadweight testers can greatly enhance a lab?s ability to calibrate wide pressure ranges with the best accuracy. With transducers and devices under test (DUTs) that have a wide measuring range, calibration can be a lengthy process, as you must change pistons in between test points so that you can calibrate its entire range.
However, a dual piston deadweight tester is equipped with a higher range piston and low range piston, that allows the technician more flexibility and efficiency when it comes to calibration. This advancement comes from decades of innovation and customer interaction to locate a solution for industry needs.
Formula for pressure calculations
Since the Industrial Revolution, high accuracy pressure measurement is a key requirement in process industries. Pressure measuring instruments like gauges were employed to measure and monitor pressure conditons. They must be calibrated against instruments of an increased amount of accuracy. Deadweight testers (generally known as pressure balances) are high accuracy instruments used in the calibration of electronic or mechanical pressure instruments. The core element of a deadweight tester is really a precisely manufactured piston-cylinder system, onto which a mass load is applied so that you can generate the average person test points in a calibration.
The mass load is proportional to the prospective pressure, which is achieved through optimally graduated masses. The design of the entire calibration system, in addition to the precise manufacturing of the piston and cylinder, ensure quality performance with long free-rotation time and low sink rates of the piston. With measurement uncertainties right down to 0.002% of reading, deadweight testers will be the golden standard in calibration.
Belief includes a single piston-cylinder system. This requires mass sets to be changed so as to calibrate multiple or wide measuring ranges. This may become a cumbersome process, one that requires many resources. In 1953, the entity today referred to as DH-Budenberg pioneered the first dual range piston-cylinder technology with two pistons housed concentrically in one unit. DH-Budenberg was later acquired by WIKA in 2011. The initial design included 1/8 in2 (low pressure) and 1/180 in2 (high pressure) pistons housed in one unit. Using a single mass-set it was now possible to create two points of equilibrium at 55 bar (low pressure) and 550 bar (high pressure) respectively. This is followed with the release of high pressure variant that could generate pressure upto 1100 bar.
Explanation dual piston design
The initial dual piston design was largely constructed of tool roll, and had an accuracy of 0.03 % of the pressure being measured in the standard model and 0.05 % in the ruthless model. In dual piston calibration, the strain on the low pressure piston is used in the auxiliary piston, that is directly linked to the weight carrier. When correct pressure is obtained, the piston head skirt floats within the low range or ?blue band.? To reach high pressure, the pressure is increased and the low pressure piston rises until a knife-edge on the bottom flange makes a seal contrary to the low pressure cylinder and acts because the high pressure cylinder. Since the low pressure piston does not contribute to the entire weight of the piston and weight carrier, the weight carrier accumulates the auxiliary weight to compensate. When correct pressure is achieved, the piston head skirt floats within the high range or ?red band.?
This iconic, patented DH-Budenberg red and blue band piston-cylinder system was re-engineered later with both piston and cylinder components manufactured from tungsten carbide. With the refinements in pressure generation and piston manufacturing processes, dual ranges are now available up to 1,400 bar with accuracies right down to 0.006 % of reading.
These developments in dual piston technology made deadweight testers, such as for example DH-Budenberg?s CPB5800 Hydraulic Pressure Balance, the premier deadweight tester for industrial applications. Naked is especially ideal for calibrating a transducer with a broad measuring range, while maintaining the high accuracy for testing and calibration. By using only a single mass set and calibrating instrument, this technique can automatically switch between multiple test points of high and low pressure, making pressure calibration seamless, efficient and economical.
Note
Further information on our pressure balances are available on the WIKA website.